4UHV Ion Pump Controller

4UHV Ion Pump Controller

The latest introduction to the VacIon family of pumps from Agilent, the 4UHV Ion Pump Controller affords you the ability to power, control and monitor independently up to 4 ion pumps, with capacities ranging from 20 to 500 liters per second. With the new low-electrical-noise design, you are afforded the freedom to eliminate or shrink electrical noise filters in critical applications. As with all Agilent vacuum products, we are dedicated to your satisfaction by offering you a complete range of service and support options to satisfy your specific needs.

Features

  • Optimized pumping speed due to auto regulation of operating voltage by the controller
  • Convenient access to all elements of the vacuum system by utilizing either the analog or RS232/485 ports
  • Multiple configurations available to suit your needs - whether operating one to four pumps with varying sizes, from 20 to 500 liters/second
  • Applied high voltage is optimized as a function of pressure, providing more accurate pressure readings, allowing pressure measurement in the 10-10 mbar range - the UHV region
  • Protect mode limits the current to protect the pump and controller.
  • Cable is equipped with an interlock system which immediately shuts down the high voltage when the plug is removed from the pump
  • The AC component of the HV output is reduced to a minimum, virtually eliminating the need for additional filters - especially useful for SEM applications
  • Utilizes Agilent's T-plus Software
  •  

Technical Specifications

Input Voltage 100 – 240 Vac (+/-10%)
Input Frequency 50/60 Hz
Dimensions 400.5 x 211.4 x 177.0 (l x w x h)
Display 4 rows with 20 characters
Available configurations 1x120W, 1x200W, 2x80W, 2x200W, 4x80W, 2x80W+1x200W
Minimum Configuration One HV card with 120W, 200W or 2x80W
Output Voltage (Open Circuit) 3, 5 and 7 kV
Output Current (Short Circuit) 40 mA for 80 W channel, 100 mA for 200 W channel
Modes of Operation Local / Serial / Remote
Front Panel Readings Voltage, Pressure, Current, Status
Safety Marks CE, C_CSA_US
Current Measurement Range 10 nA to 100 mA
Input Signals On/off; Protect; Step Mode;
Output Signals Analog Out; NC Set-point; NO Set-point
HV Connector Fischer Type 105
Output Power Maximum 400 W

 

Documents Download

【返回】